Directorate G
Unit 4 - Pressure Equipment, Medical Devices, Metrology
MEDICAL DEVICES : Guidance document
MEDDEV 2. 1/5
June 1998
GUIDELINES RELATING TO THE APPLICATION OF :
THE COUNCIL DIRECTIVE 90/385/EEC ON ACTIVE IMPLANTABLE MEDICAL DEVICES
THE COUNCIL DIRECTIVE 93/42/EEC ON MEDICAL DEVICES
MEDICAL DEVICES WITH A MEASURING FUNCTION
--- (())---
Background
Annex VII, paragraph 5 of MDD requires for class I devices with a measuring functionthat the manufacturer must also follow one of the procedures referred to in annex IV, Vor VI, for the «aspects of manufacture concerned with the conformity of the productswith the metrological requirements ».
It is therefore necessary to specify criteria for the existence of a « measuring function »in a medical device.
Criteria for devices with a measuring function
The following criteria, if fulfilled together, indicate that a device has a measuringfunction:a)
The device is intended by the manufacturer to measure:--b)
quantitatively a physiological or anatomical parameter, or
a quantity or a qualifiable characteristic of energy or of substancesdelivered to or removed from the human body.
The result of the measurement--is displayed in legal units or other acceptable units within the meaning ofDirective 80/181/ECC1 or
is compared to at least one point of reference indicated in legal units orother acceptable units in compliance with the pre-mentioned directive.
c)
The intended purpose implies accuracy, claimed explicitly or implicitly, where anon-compliance with the implied accuracy could result in a significant adverseeffect on the patient’s health and safety.
Note 1:
The expression « claimed implicitly » covers cases where the user, on thebasis of the designation of the device or of its accompanying documents,or on the basis of the common use is entitled to expect accuracy where theaccuracy of the measurement has an impact on the diagnosis or therapy ofthe patient.
Measuring activities during the manufacturing process including those forcalibration purposes are not covered by this recommendation and do notimply a measuring function of the manufactured device.
Note 2:
Examples for devices with a measuring function-device for measuring body temperature, 1 OJ n° L39/40, 15.2.1980 as amended by Directive 89/617/EEC, OJ n° L357/28, 7.12.1989
2
-----
pacifier which includes a temperature display including those with only a changeof colour where criteria b is met,
device for indicating that a body temperature is above or below a specified value,non-active non-invasive device for measuring blood pressure ,non-active device for measuring intra-ocular pressure,
device for measuring volume or pressure or flow of liquid or gases delivered to orremoved from the human body (included any container with a graduation scale orwith a single point graduation where criteria c is met).
Examples for devices without a measuring function----patch for indicating trends of body temperature (where criteria b is not met),device for the delivery of liquid to the human body (e.g. medicine spoons, cups,droppers, without graduation or scale or display of measuring unit),
device for displaying trends of physiological parameters (e.g. urine bags withoutgraduation or scale, callipers for obesity),eye-test chart.